MAQUET SAS POWERLED; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number PWD700DF AIM K3
Device Problem Detachment Of Device Component (1104)
Patient Problem No Patient Involvement (2645)
Event Date 04/20/2016
Event Type  malfunction  
Manufacturer Narrative
Maquet was not able to evaluate the device, nor was able to evaluate the ambient light that failed.Based on the device serial number, maquet is able to determine that the issue is due to the breakage of the fixing tabs.The biomedical technician replaced the ambient sector with a new one of a different design.
 
Event Description
The customer reported that an ambient light is falling out, dangling still attached via the cable.No patient was involved at the moment of the incident and no injury was reported.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERLED
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
orléans cedex 2
FR 
Manufacturer (Section G)
MARIE-FRANÇOISE CABEL - MAQUET SAS
parc de limère
avenue de la pomme de pi
orléans cedex 2 45074
FR   45074
Manufacturer Contact
parc de limère
avenue de la pomme de pi
orléans cedex 2 45074
MDR Report Key5673045
MDR Text Key46289538
Report Number9710055-2016-00036
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
PMA/PMN Number
K070442
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Remedial Action Repair
Type of Report Initial
Report Date 04/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPWD700DF AIM K3
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/20/2016
Date Device Manufactured12/05/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1730-2014
Patient Sequence Number1
-
-