• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMEDA, INC. PURELY YOURS; ELECTRIC BREAST PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMEDA, INC. PURELY YOURS; ELECTRIC BREAST PUMP Back to Search Results
Model Number 24502082
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/23/2016
Event Type  malfunction  
Manufacturer Narrative
Customer was emailed and phoned twice requesting return of the purely yours breast pump base to ameda, inc.For investigation of alleged leaking fluid.As of this date, 05/23/2016, the pump base has not been returned to ameda, therefore no visual inspection or testing could be conducted.
 
Event Description
Customer contacted ameda, inc.On (b)(6) 2016 to report batteries exploded and leaked battery fluid inside battery compartment of her purely yours breast pump during usage on (b)(6) 2016.Customer was pumping while traveling in the car when this event occurred.She denies any battery fluid leaking onto her therefore no injury or burn occurred.Replacement purely yours breast pump was shipped overnight to customer.
 
Manufacturer Narrative
The returned product was tested per ameda engineering protocol to determine whether the allegation of leaking fluid could be confirmed.The returned product was assessed for indications of malfunction or thermal event.The returned product was assessed for functionality and met functional specifications.Dark grey substance, e.G battery acid, was observed inside the battery compartment.Internally, no signs of foreign substance, burning, melting or charring were observed.Additional testing confirmed that batteries may leak when the upper middle battery is placed incorrectly, with the positive and negative end reversed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PURELY YOURS
Type of Device
ELECTRIC BREAST PUMP
Manufacturer (Section D)
AMEDA, INC.
485 half day rd.
suite 320
buffalo grove IL 60089
Manufacturer Contact
linda zager
485 half day rd.
suite 320
buffalo grove, IL 60089
8479642620
MDR Report Key5673212
MDR Text Key45699690
Report Number3009974348-2016-00207
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973501
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Remedial Action Replace
Type of Report Initial,Followup
Report Date 04/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number24502082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received04/25/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/12/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age31 YR
-
-