The device has not been returned for evaluation; without return of the device, no definitive conclusions can be drawn regarding the clinical observation.Should the device be returned or additional information become available, a supplemental report will be submitted.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
Medtronic received information that during treatment of an aneurysm located in the ophthalmic segment of the internal carotid artery, the "head" of the pipeline device was fixed in the capture coil and could not be released.The push wire was rotated several times, but not more than the recommended 10 times.The physician used a new device to complete the procedure successfully.No patient injury was reported.
|
Suspect medical device: added lot number the pipeline pushwire and braid were returned for evaluation.The tip coil was observed to be stretched, and the capture coil was observed to be damaged.The pipeline braid was returned detached from the pushwire; therefore, the distal and proximal ends of the braid were unable to be identified.The pipeline braid was observed to be fully open with both ends having slight fraying, and the middle section having no damage.Based on the analysis findings, the clinical observation could not be confirmed.We are unable to definitively determine the cause of the reported event or damages observed.Per the instructions for use: ¿if the capture coil tip of the delivery system becomes stuck in the mesh of a delivered ped, rotate the wire clockwise while advancing the wire to try to release it, then slowly pull back on the delivery wire.A review of this device lot history record was conducted and no issues were identified that could have contributed to this event.All products are 100% inspected for damage and irregularities during manufacture.
|