Halyard received a single report that referenced three different incidences, which were associated with separate units, involving no patient contact.
This is the third of three reports.
Refer to 1054380-2016-00002 for the first report.
Refer to 1054380-2016-00003 for the second report.
It was reported by the customer that, there is an issue with tearing on the sterilization wrap.
This incident occurred post sterilization with no patient or instrument contact.
No additional information was provided.
|
The actual complaint product was returned for evaluation.
Two used samples were received with no packaging.
The samples have sterilization indicator tape affixed to the outer layer.
Sample a: the customer has circled one area on the sample.
With visual inspection, no tears or holes are seen in the circled area.
There is a visible pressure point seen, but it has not created a hole or tear.
The area was then viewed under magnification.
No holes or tears are seen under magnification.
The reported failure is not confirmed based on the assessment of the circled area.
Sample 2b: the customer has circled one area on the sample.
With visual inspection, no tears or holes are seen in the circled area.
There is a visible pressure point seen, but it has not created a hole or tear.
The area was then viewed under magnification.
No holes or tears are seen under magnification.
The reported failure is not confirmed based on the assessment of the circled area.
Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.
This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
|