• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY MICROSTAAR® INJECTOR; INTRAOCULAR LENS GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY MICROSTAAR® INJECTOR; INTRAOCULAR LENS GUIDE Back to Search Results
Model Number MSI-TM
Device Problems Scratched Material (3020); Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Pt age and weight: unk.Event date: unk.Implant and explant dates: n/a.Device evaluated by manufacturer? no.The product for this complaint was not returned.However, the lens was returned and visual inspection found a piece missing from the lens optic and one haptic bent.The lens was returned dry and there was evidence of surgical residue.Event problem and evaluation codes: (b)(4).Injector not returned.
 
Event Description
The reporter stated the injector overrode and scratched the optic of the aq5010v three-piece silicone lens, -1.0 diopter, as the surgeon loaded the lens into the injector.There was no patient contact.The backup lens was implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROSTAAR® INJECTOR
Type of Device
INTRAOCULAR LENS GUIDE
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
michelle andres
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key5685229
MDR Text Key46095357
Report Number2023826-2016-00769
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K940593
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/27/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date11/30/2018
Device Model NumberMSI-TM
Device Catalogue NumberN/A
Device Lot Number1346218
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/28/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/15/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
LENS MODEL: AQ5010V - SERIAL #(B)(4)
-
-