Catalog Number IGTCFS-65-FEM-CELECT-PERM |
Device Problems
Complete Blockage (1094); Difficult to Remove (1528); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Occlusion (1984); Pain (1994); Swelling (2091); Ulceration (2116); Injury (2348)
|
Event Date 09/04/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Lot#: unknown as information was not provided.Catalog#: unknown but referred to as a cook celect filter.Expiration date: unknown as lot# is unknown.Since catalog# is unknown the 510(k) could be either k061815, k073374 or k090140.Investigation is still in progress.
|
|
Event Description
|
Description according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2010 at (b)(6)." patient outcome: it is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
|
|
Manufacturer Narrative
|
(b)(4).Igtcfs-65-fem-celect-perm.Investigation is still in progress.
|
|
Event Description
|
Description according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2010 at (b)(6)." patient outcome: it is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
|
|
Manufacturer Narrative
|
(b)(4).Corrected data based on new information received: adverse event to product problem, serious injury to malfunction.The event is currently under investigation.A supplemental report will be provided upon conclusion.
|
|
Event Description
|
This additional information was received on 06/14/2016 as follows: the plaintiff allegedly received the filter implant on (b)(6) 2010 due to dvt.An unsuccessful retrieval attempt allegedly occurred on (b)(6) 2015.The plaintiff alleges clogged filter, device cannot be retrieved, pain and swelling in both legs, ulceration in right leg, and anxiety.
|
|
Manufacturer Narrative
|
Manufacturer ref# (b)(4).Exemption number e2016032.(b)(4).It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'clogged filter; device cannot be retrieved; pain and swelling; ulceration; anxiety'.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.Ivc thrombotic occlusion as an outcome for cook ivc filters is addressed in the published scientific literature.Ivc thrombotic occlusion is defined as the presence of an occluding thrombus in the ivc after filter insertion and documented by ultrasound (us), ct, mr imaging or venography; this may be symptomatic or asymptomatic.Unknown if the reported pain, swelling, and anxiety is directly related to the filter and unable to identify a corresponding failure mode at this time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
|
|
Event Description
|
No additional information to report at this time.
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unchanged, unknown or unavailable.Exemption number (b)(4).William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Name and address for importer site: (b)(4).Investigation is reopened due to additional information provided.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating "celect: thrombotic occlusion, unable to retrieve, pain, swelling, ulceration, anxiety - updated sfc." cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.Ivc thrombotic occlusion as an outcome for cook ivc filters is addressed in the published scientific literature.Ivc thrombotic occlusion is defined as the presence of an occluding thrombus in the ivc after filter insertion and documented by ultrasound (us), ct, mr imaging or venography; this may be symptomatic or asymptomatic.Unknown if the reported pain, swelling, ulceration and anxiety is directly related to the filter and unable to identify a corresponding failure mode at this time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Manufacturer Narrative
|
Exemption number (b)(4).William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Name and address for importer site: (b)(4).Blank fields on this form indicate the information is unknown or unavailable, or unchanged.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Per an angiogram performed on (b)(6) 2015 there is 'complete thrombotic occlusion extending from the popliteal vein in the left lower extremity into the inferior vena cava to the level of a pre-existing ivc filter'.
|
|
Search Alerts/Recalls
|