ROCHE DIAGNOSTICS URISYS 1100; AUTOMATED URINE ANALYZER Back to Search Results
Model Number U1100
Device Problem Low Test Results (2458)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/31/2016
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer received questionable urinalysis results for one patient sample.Of the data provided, only the results for glucose were discrepant.The glucose result read visually was 2+.The result from the urisys 1100 was negative.It was unknown if any erroneous result was reported outside the laboratory.The patient was not adversely affected.The lot number and expiration date of the urine strips was requested, but was not provided.
 
Manufacturer Narrative
A specific root cause could not be identified.Additional information for further investigation was requested but was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URISYS 1100
Type of Device
AUTOMATED URINE ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5701548
MDR Text Key46669359
Report Number1823260-2016-00731
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K033548
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Medical Technologist
Type of Report Initial,Followup
Report Date 07/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberU1100
Device Catalogue Number03617548001
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/31/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-