ALERE SAN DIEGO, INC. INRATIO PT/INR TEST STRIPS; PROTHROMBIN TIME TEST Back to Search Results
Model Number 100071
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Improper or Incorrect Procedure or Method (2017); Incorrect Or Inadequate Test Results (2456)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 05/12/2016
Event Type  malfunction  
Manufacturer Narrative
Investigation pending.
 
Event Description
A variance was reported between inratio inr results.The results were as follows: on (b)(6) 2016: inratio=4.3, inratio=1.6 , 5 min apart.On (b)(6) 2016: inratio=1.7 , 20 mins after 1.6 result.Reported therapeutic range: 2-3.It was reported that the finger was being milked after the finger stick and multiple finger sticks were performed on the same finger.For the inratio=1.7, a new finger was used for testing.
 
Manufacturer Narrative
The product was not returned for evaluation.Therefore, a review of the entire testing history for the reported lot was performed.In-house testing on strip lot 370105ar meets release criteria.The product performed as expected.A review of the manufacturing records for the lot did not uncover any relevant non-conformances; the lot met release specifications.Technique issues were identified in the complaint which may have contributed to the precision issue experienced by the customer.Based on the information available, there is no indication of a product deficiency.No corrective action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INRATIO PT/INR TEST STRIPS
Type of Device
PROTHROMBIN TIME TEST
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9975 summers ridge rd.
san diego CA 92121
Manufacturer (Section G)
ALERE SAN DIEGO, INC.
9975 summers ridge rd.
san diego CA 92121
Manufacturer Contact
ya-ling king
9975 summers ridge rd.
san diego, CA 92121
8588052084
MDR Report Key5704855
MDR Text Key48030003
Report Number2027969-2016-00425
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092987
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number100071
Device Lot Number370105AR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/09/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-