ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 05/13/2016
Event Type  malfunction  
Manufacturer Narrative
The autopulse platform (s/n (b)(4)) was returned to zoll for evaluation.Investigation results as follows: device history record (dhr) was reviewed for service information related to the reported complaint and there was no previous history of complaint reported for autopulse with serial number (b)(4).Visual inspection of the returned platform was performed and found no physical damages to the autopulse platform.A review of the archive was performed and no discrepancies were observed.During functional testing the autopulse platform passed all functional tests without any discrepancies.Further testing showed that the backlight on the platform was not working and was attributed to a faulty processor board.In summary the customer's reported complaint is confirmed.The root cause for the display not to be clear and having intermittent backlight was due to a faulty processor board.After the part was replaced, the platform passed all final functional testing criteria.
 
Event Description
The customer reported that during shift check, the backlight on the autopulse platform was intermittently not working.The display was not clearly readable.There was no patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
joy patel
2000 ringwood ave,
san jose, CA 95131
4084192957
MDR Report Key5705284
MDR Text Key46760854
Report Number3010617000-2016-00395
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/13/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/27/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-