• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE INSULIN INFUSION PUMP Back to Search Results
Device Problem Device Stops Intermittently (1599)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The pump has not been returned to animas for evaluation. If the device is returned, an evaluation shall be completed and a supplemental report will be filed. No conclusions can be made at this time. Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
 
Event Description
On (b)(6) 2016, the reporter contacted animas, alleging a power (intermittent power) issue. This complaint is being reported because the reported issue was not resolved with troubleshooting. There was no indication that the product caused or contributed to an adverse event.
 
Manufacturer Narrative
Follow-up #1: date of submission 07/21/2016. Device evaluation: the device has been returned and evaluated by product analysis on 07/05/2016 with the following findings: the black box shows no evidence of por events. The battery compartment and the battery cap are undamaged and able to fit securely, cap contact height and width measurements are within specifications. The battery cap was fastened and then unscrewed ½ turn with no reboots occurring. The ¿ez-prime¿ steps were performed correctly, no power interruption occurred during the investigation, the product performs within specifications. The pump¿s cover was removed, no intermittent condition was found to the pcb or to the power flex. Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameANIMAS VIBE
Type of DeviceINSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key5705286
MDR Text Key47406018
Report Number2531779-2016-11284
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 05/23/2016
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Device Age1 MO
Event Location No Information
Date Manufacturer Received05/23/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/07/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-