PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 04/26/2016
Event Type  malfunction  
Manufacturer Narrative
This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.Product event summary: analysis found that the device failed functional testing for over voltage current drain testing.The device was repaired making it compatible with all batteries.It was also noted that the serial number label was damaged.
 
Event Description
It was reported that the device was defective, however no further details were available.The external pulse generator (epg) was returned for servicing.There was no patient involvement.
 
Manufacturer Narrative
Further review prompted a change in the device analysis results.Investigation noted that there was no device failure during testing, but rather an issue with the tester itself.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5712869
MDR Text Key47155565
Report Number3004593495-2016-00219
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 04/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received04/26/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/07/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1356-2015
Patient Sequence Number1
-
-