MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AESPIRE 7100; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AESPIRE 7100; ANESTHESIA GAS MACHINE Back to Search Results
Device Problems Electrical /Electronic Property Problem (1198); Arcing (2583)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/10/2016
Event Type  malfunction  
Manufacturer Narrative
Patient information could not be obtained due to country privacy laws.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Date of device manufacture year is 2007.The month of manufacture was unavailable at time of mdr filing.A ge healthcare service representative performed a checkout of the equipment and confirmed the reported complaint.The display power supply cord and 2 fuses were replaced.The unit was returned to service.
 
Event Description
The hospital reported that an electrical arc that occurred below the display caused the unit to shutdown unexpectedly during a case.The unit was swapped out to complete the case.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESPIRE 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda drive
madison WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda drive
madison WI 53718
Manufacturer Contact
stephanie cass
3000 n. grandview boulevard
waukesha, WI 53188
MDR Report Key5717510
MDR Text Key47207976
Report Number2112667-2016-01107
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K000706
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 06/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/1970
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-