SORIN GROUP DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 25-40-45
Device Problem Insufficient Flow or Under Infusion (2182)
Patient Problem No Patient Involvement (2645)
Event Date 05/17/2016
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Sorin group (b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the s5 gas blender system experienced low flow during maintenance.There was no patient involvement.The service representative requested that the device be returned to sorin group deutschland for further investigation.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the s5 gas blender system experienced low flow during maintenance.There was no patient involvement.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the s5 gas blender system.The incident occured in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).The complained gas blender was returned to sorin group (b)(4) for investigation.The reported issue was confirmed during functional testing and was traced to a defective air valve.The air valve was replaced and a functional check and new calibration were performed.Functional control and a technical safety inspection were successfully carried out and no further issues were discovered.The device was cleaned and disinfected and returned to the customer.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Sorin group (b)(4) will continue to monitor for trends related to this type of issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 GAS BLENDER SYSTEM
Type of Device
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key5718824
MDR Text Key47201974
Report Number9611109-2016-00339
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K101046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 05/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25-40-45
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/20/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-