• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TIBIAL ALIGNMENT ANKLE CLAMP EM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TIBIAL ALIGNMENT ANKLE CLAMP EM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 6541-2-609
Device Problems Break (1069); Fracture (1260); Nonstandard Device (1420); Material Integrity Problem (2978)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 05/17/2016
Event Type  malfunction  
Manufacturer Narrative
When completed, the investigation results will be submitted in a supplemental report.
 
Event Description
During incoming inspection of the device it was noticed that the item was damaged.
 
Manufacturer Narrative
A device history review indicates devices were manufactured and approved into final stock with no reported discrepancies.The complaint history review indicates there has been one other event for the reported lot.The reported event was confirmed.One of the flippers of the tibial alignment ankle clamp was fractured.The device was manufactured to revision g of the product drawing.The damaged device was discovered during inspection; there was no surgical procedure associated with the reported event.This event meets the definition of preventive maintenance; no further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
During incoming inspection of the device it was noticed that the item was damaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL ALIGNMENT ANKLE CLAMP EM
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5721028
MDR Text Key48165206
Report Number0002249697-2016-01937
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123486
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 05/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number6541-2-609
Device Lot NumberN4S26
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received07/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/04/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number2249697-7/12/2013-008C
Patient Sequence Number1
Patient Outcome(s) Other;
-
-