It was reported that there was an inability to get a reading for the generator¿s heartbeat detection during implant surgery.Troubleshooting was attempted, including attempting different sensitivity values and moving the generator closer to the heart, but did not succeed.The case was completed with the auto-stimulation mode disabled.Follow up revealed that the pre-surgical evaluation to establish candidacy for the auto-stimulation mode was not performed.During the surgery, the current modes were not disabled.Lead impedance was within normal limits.A review of programming history from the surgery showed that sensitivity values 3, 4, and 5 were attempted.As there were no device interrogations or diagnostics performed while the heartbeat detection was enabled, it is not expected that review of the interval device data would yield pertinent information at this time.The generator device history record was reviewed and found all specifications were met prior to distribution, including the r-wave detection functional testing.
|