It was reported that the intended m106 replacement generator was unable to detect the heart rate within surgery.A reading of ¿?????¿ was present in the heart rate field on all sensitivity settings.No pre-surgical evaluation was performed.Device diagnostics were within normal limits.The programmed currents were turned off during attempts to verify heart rate.All available sensitivity values were cycled through and the issue persisted.Recommendations were given to mitigate emi, to reposition in the pocket, and reposition the wand.A different generator was ultimately implanted without reported issue.The generator device history record was reviewed and found all specifications met prior to distribution.In particular, all functional specifications, including r-wave verification, were completed prior to distribution.The internal device data of the generator was reviewed.The three available lines of interrogation and diagnostics with the tachycardia detection enabled showed that a sensitivity setting of 5 was used.The detected foreground heart rate values were 81.2, 81.2, and 118.7 bpm.There were no anomalies with impedance or battery voltage noted.The suspect generator was received by the manufacturer on 06/13/2016 and is undergoing product analysis.
|