Catalog Number 72200616
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/18/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during a knee arthroscopy procedure, the device would not surpass 500 rpm.The procedure was completed with a competitor's device.No patient injury or complications were reported.
 
Manufacturer Narrative
A visual inspection was performed on the product and no damage was observed.Complaint of low rpms was confirmed.Mdu failed for hand piece blade stall error as well as overheating.The motor/gearbox was removed from the housing.The gearbox was removed from motor and motor tested good.The gearbox appears to be corroded internally.The root cause of this event was identified to be associated with corrosion of the gearbox assembly.A motor stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing.Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved.Udi# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
james gonzales
150 minuteman road
andover, MA 01810
5123585706
MDR Report Key5739575
MDR Text Key48665854
Report Number1643264-2016-00097
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received08/16/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/31/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-