• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION ALARIS® PUMP MODULE ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION ALARIS® PUMP MODULE ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 2420-0007
Device Problems Stretched (1601); Deformation Due to Compressive Stress (2889); Material Distortion (2977)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
 
Event Description
The customer reported ballooning of the silicone segment of the tubing, which caused alarms and error codes on the pump.The tubing was replaced; there was no harm reported.
 
Manufacturer Narrative
The customer¿s report of a bulge/balloon in the silicone segment below the upper fitment was confirmed.Visual inspection showed that the silicone segment tubing had a slight bulge, discoloration, and was weakened near its upper portion just below the upper fitment.Functional testing was unable to replicate the ballooning.The root cause for the reported bulge/balloon in the tubing could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS® PUMP MODULE ADMINISTRATION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
stephen bilello
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key5756876
MDR Text Key48402652
Report Number9616066-2016-00842
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/28/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/01/2019
Device Model Number2420-0007
Device Catalogue Number2420-0007
Device Lot Number16047293
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received08/08/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/25/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-