STRYKER ORTHOPAEDICS-MAHWAH SPECIALTY TRIATHLON PS TIBIAL INSERT TRIAL SIZE 3 SET; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number I-K3497PS03
Device Problems Break (1069); Fracture (1260); Material Integrity Problem (2978)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 06/06/2016
Event Type  malfunction  
Manufacturer Narrative
When completed, the investigation results will be submitted in a supplemental report.
 
Event Description
When surgeon removed trial from tibial tray, the trial broke into pieces.Surgeon was able to recover both.
 
Manufacturer Narrative
An event regarding crack/fracture involving an specialty trial was reported.The event was confirmed.The device was returned in two pieces confirming the event.It exhibited indentation marks consistent with normal use.Examination with material analysis team member confirmed the fracture of the returned device apparently from multiple overload conditions.No medical records or x-rays were made available for evaluation.Device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies there has been no other event for the lot referenced.The device was returned in two pieces confirming the event.Examination with material analysis team member confirmed the fracture of the returned device apparently from multiple overload conditions.If additional information are received, this investigation will be reopened and re-evaluated.
 
Event Description
When surgeon removed trial from tibial tray, the trial broke into pieces.Surgeon was able to recover both.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECIALTY TRIATHLON PS TIBIAL INSERT TRIAL SIZE 3 SET
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5761452
MDR Text Key49185985
Report Number0002249697-2016-02137
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberI-K3497PS03
Device Lot Number18TCF
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/30/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/25/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-