STRYKER ORTHOPAEDICS-MAHWAH SPECIALTY TRIATHLON PS TIBIAL INSERT TRIAL SIZE 3 SET; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
Catalog Number I-K3497PS03 |
Device Problems
Break (1069); Fracture (1260); Material Integrity Problem (2978)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/06/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
When completed, the investigation results will be submitted in a supplemental report.
|
|
Event Description
|
When surgeon removed trial from tibial tray, the trial broke into pieces.Surgeon was able to recover both.
|
|
Manufacturer Narrative
|
An event regarding crack/fracture involving an specialty trial was reported.The event was confirmed.The device was returned in two pieces confirming the event.It exhibited indentation marks consistent with normal use.Examination with material analysis team member confirmed the fracture of the returned device apparently from multiple overload conditions.No medical records or x-rays were made available for evaluation.Device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies there has been no other event for the lot referenced.The device was returned in two pieces confirming the event.Examination with material analysis team member confirmed the fracture of the returned device apparently from multiple overload conditions.If additional information are received, this investigation will be reopened and re-evaluated.
|
|
Event Description
|
When surgeon removed trial from tibial tray, the trial broke into pieces.Surgeon was able to recover both.
|
|