• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The pump has not been returned to animas at the time of this report.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event Description
On (b)(6) 2016, a reporter contacted animas alleging that the display screen froze.This complaint is being reported because the reported issue was not resolved through troubleshooting.There was no allegation of an adverse event associated with this complaint.
 
Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on (b)(6) 2017 with the following findings:.Unable to investigate display issue complaint due to keypad defect.(b)(4).Opened pump, moisture damage found on keypad flex connector.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key5767084
MDR Text Key49283715
Report Number2531779-2016-14429
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 06/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Device Age13 MO
Date Manufacturer Received06/18/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age13 YR
-
-