Patient underwent explant due to pain as reported in mfr.Report # 1644487-2016-01153.An analysis was performed on the returned lead portions.The electrodes portion of the leads was not returned for analysis and therefore a complete evaluation could not be performed on the entire lead product.Abraded openings were found on the outer and inner silicone tubing.These abraded openings and the cut ends that were made during the explanted process most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.What appeared to be white deposits were observed in various locations.Eds (energy dispersion spectroscopy ¿ provides chemical or element identity/composition analysis) was performed on the deposit observed on the marked connector boot and identified the deposit as containing silicon, phosphorus, sodium, magnesium, sulphur and calcium.Refer to attached eds sheet for additional information.With the exception of the inner tubing abraded opening, the condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.The setscrew marks found on the connector pins provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, with no discontinuities identified.Based on the findings, there is evidence of an inner tubing abraded opening.
|