The investigation determined that lower than expected tsh results were obtained from two vitros quality control fluids and two patient samples using vitros immunodiagnostics products tsh reagent in combination with a vitros 5600 integrated system.The most likely assignable cause of the lower than expected qc results is the use of an expired calibration by the user.The results detailed above were obtained on (b)(6) 2016 using a calibration from (b)(6) 2015.Per the vitros tsh instructions for use the assay should be calibrated at least every 28 days.Additionally, insufficient quality control data was available for analysis to completely rule out a reagent performance issue and precision testing was not performed so unexpected instrument performance can also not be entirely ruled out as contributing to the events.The customer recalibrated their instrument and obtained acceptable qc results and was consequently satisfied with the performance of the tsh reagent.
|