BIOTRONIK SE & CO. KG EVIA HF-T; CRT-P Back to Search Results
Model Number 377208
Device Problem Premature Elective Replacement Indicator (1483)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/25/2016
Event Type  malfunction  
Event Description
This device was at eri indication and replaced with an oem device.Should additional information be received, this file will be updated.
 
Manufacturer Narrative
The pacemaker was returned and analyzed.The memory content demonstrated a normal functionality of the device while implanted and in service.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.There was no indication of a device malfunction.In agreement with the clinical observation, the device interrogation revealed the battery status eri.The archive data and the programmed parameters were inspected.High left ventricular pacing outputs of 5.6 v and 1.5 ms were documented in the archive data.This represents a high current program, which results in a faster discharge of the battery.The amount of charge taken from the battery was verified and the battery condition was found to be as expected.In conclusion, the pacemaker was fully functional.The battery status was anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIA HF-T
Type of Device
CRT-P
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin D-123 59
GM  D-12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key5781932
MDR Text Key49148231
Report Number1028232-2016-02471
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P070008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 06/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number377208
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/05/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age70 YR
-
-