Catalog Number 7206010
Device Problems Fluid/Blood Leak (1250); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/17/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that black fluid was leaking out of the 5.5mm full radius blade and into the patient's joint.A backup was used to complete the procedure.It was noted that the black fluid was removed via suction.Delay was less than 30 minutes and no patient impact was noted.
 
Manufacturer Narrative
Device investigation narrative - one 5.5 full radius blade was returned for evaluation.Visual assessment showed the device is contaminated with an unidentifiable residue.Examination of the inner and outer blade surfaces showed no evidence of debridement.The condition of the blade eliminates it as the source of the reported contamination.Further investigation is not warranted at this time.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FULL RADIUS BLADE,5.5MM,SER 3000
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
james gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key5784289
MDR Text Key49906380
Report Number1219602-2016-00368
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/07/2018
Device Catalogue Number7206010
Device Lot Number50595113
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received09/08/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/07/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-