• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND VITALITY; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND VITALITY; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number T180
Device Problems Material Integrity Problem (2978); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Endocarditis (1834); Unspecified Infection (1930)
Event Date 04/27/2016
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that the patient was admitted to the hospital for severe endocarditis and scheduled for a device replacement procedure.During the extraction the physician attempted to remove the device from the pocket by grasping it with a pair of forcepts and pulling.After multiple attempts, the header began to separate from the battery portion of the device.Eventually the header completely separated from the battery.All parts of the device and lead were removed without further incident.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALITY
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5789161
MDR Text Key49366953
Report Number2124215-2016-07803
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/22/2008
Device Model NumberT180
Other Device ID NumberVITALITY DR HE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/27/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/24/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
(B)(4)
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age56 YR
-
-