COOK ENDOSCOPY HERCULES 3 STAGE BALLOON ESOPHAGEAL KNQ, DILATOR, ESOPHAGEAL Back to Search Results
Catalog Number HBD-18-19-20
Device Problems Hole In Material; Use of Device Problem
Event Type  Malfunction  
Manufacturer Narrative

Concomitant products: boston scientific alliance inflation handle. Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation. The report could not be confirmed. A review of the device history record could not be conducted because the lot number was not provided. Investigation conclusions: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation. A definitive cause for the reported observation could not be determined. The reporter stated the balloon was not lubricated prior to advancement through the accessory channel of the endoscope. A possible contributing factor to balloon damage is inadequate lubrication of the balloon with a lubricating agent. The instructions for use direct the user to "apply a lubricating agent to the balloon to facilitate passage through the endoscope accessory channel. " this activity will aid in endoscopic advancement and balloon preservation. The reporter stated negative pressure was not applied to the balloon device prior to advancement through the accessory channel of the endoscope. Another possible contributing factor to balloon damage is failure to apply negative pressure to the balloon dilator prior to advancement through the endoscope. The instructions for use direct the user "to facilitate passage through the endoscope, apply negative pressure to the catheter. " the application of negative pressure will aspirate all residual air from the balloon and ease endoscopic advancement. Negative pressure will also aid in balloon preservation and optimize balloon performance. Damage to the balloon material can occur if the balloon is inflated prior to advancement through the endoscope or if the balloon is inflated while partially or fully inside the accessory channel of the endoscope. The instructions for use contain the following precaution: ¿do not pre-inflate the balloon. ¿ the instructions for use contain the following warning: ¿during dilation, do not inflate balloon beyond the maximum indicated inflation pressure. ¿ over inflation can cause damage to the balloon dilator. Another possible contributing factor is using a compromised inflation device to inflate the balloon. If the pressure reading of the inflation device is inaccurate, this could contribute to over inflation, possibly resulting in damage to the balloon material. Prior to distribution, all hercules 3 stage balloon esophageal are subjected to a visual examination to ensure device integrity. Corrective action: a review of the complaint history was conducted. The likelihood of occurrence is considered rare. Corrective action is not warranted at this time based on the quality engineering risk assessment. Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available. Additional comments regarding this report: based on the information provided, a cook representative has been directed to contact the medical facility involved in an effort to promote further education and understanding related to appropriate usage of this product.

 
Event Description

During an esophageal stricture dilation, the physician used four (4) cook hercules 3 stage balloon esophageal. The product did not fill properly and water flew out of two holes. Reference mdr number: 1037905-2016-00244.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHERCULES 3 STAGE BALLOON ESOPHAGEAL
Type of DeviceKNQ, DILATOR, ESOPHAGEAL
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem , NC 27105
3367440157
MDR Report Key5798312
Report Number1037905-2016-00245
Device Sequence Number1
Product CodeKNQ
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,FOREIG
Reporter Occupation
Type of Report Initial
Report Date 06/21/2016
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received07/15/2016
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue NumberHBD-18-19-20
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location Hospital
Date Manufacturer Received06/21/2016
Was Device Evaluated By Manufacturer? No Answer Provided
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 07/15/2016 Patient Sequence Number: 1
Treatment
FUJINON GASTROSCOPE, MODEL EG-530CT
-
-