Catalog Number HBD-W-10-11-12
Device Problem Hole In Material (1293)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/16/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This mdr report is part of the 227 pilot program.For the two (2) reported events, the devices were not returned to cook endoscopy.A product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.A review of the device history record could not be conducted because the lot numbers were not provided.We could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The reporter was unable to specify if the balloon was lubricated prior to advancement through the accessory channel of the endoscope.A possible contributing factor to balloon damage is inadequate lubrication of the balloon with a lubricating agent.The instructions for use direct the user to "apply a lubricating agent to the balloon to facilitate passage through the endoscope accessory channel." this activity will aid in endoscopic advancement and balloon preservation.Another possible contributing factor to balloon damage is failure to apply negative pressure to the balloon dilator prior to advancement through the endoscope.The instructions for use direct the user to "to facilitate passage through the endoscope, apply negative pressure to the device." the application of negative pressure will aspirate all residual air from the balloon and ease endoscopic advancement.Negative pressure will also aid in balloon preservation and optimize balloon performance.Damage to the balloon material can occur if the balloon is inflated prior to advancement through the endoscope or if the balloon is inflated while partially or fully inside the accessory channel of the endoscope.The instructions for use contain the following precaution: ¿do not pre-inflate the balloon.¿ the instructions for use contain the following warning: ¿during dilation, do not inflate balloon beyond the maximum indicated inflation pressure.¿ over inflation can cause damage to the balloon dilator.Another possible contributing factor is using a compromised inflation device to inflate the balloon.If the pressure reading of the inflation device is inaccurate, this could contribute to over inflation, possibly resulting in damage to the balloon material.Prior to distribution, all hercules 3 stage wireguided balloon esophageal-pyloric-colonic are subjected to a visual examination to ensure device integrity.A review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
This report summarizes two (2) malfunction events.A review of the events indicated that during an esophageal dilation procedure, the physician used two (2) cook hercules 3 stage wireguided balloon esophageal-pyloric-colonic.[there was a] leak in the balloons.The two (2) events involved a patient with no consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HERCULES 3 STAGE WIREGUIDED BALLOON ESOPHAGEAL-PYLORIC-COLONIC
Type of Device
KNQ, DILATOR, ESOPHAGEAL
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key5808292
MDR Text Key49970154
Report Number1037905-2016-00214
Device Sequence Number1
Product Code KNQ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported2
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHBD-W-10-11-12
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date05/16/2016
Event Location Hospital
Date Manufacturer Received05/16/2016
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OLYMPUS ENDOSCOPE (UNKNOWN MODEL)
-
-