• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS 8000 PERFUSION SYSTEM; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE - SARNS ROLLER PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS 8000 PERFUSION SYSTEM; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE - SARNS ROLLER PUMP Back to Search Results
Model Number 16402
Device Problem Naturally Worn (2988)
Patient Problem No Patient Involvement (2645)
Event Date 06/28/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This complaint is related to emdr #1828100-2015-00253.Per the srt, the indicator dial will be replaced with new part and returned for evaluation.
 
Event Description
The service repair technician (srt) reported that during routine testing of the device at the service center, the roller pump's indicator dial silks-creen increment markings are worn off.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.The service repair technician (srt) replaced the indicator dial with new part.The srt performed preventive maintenance (pm) and verification/release test.The roller pump operated to the manufacturer's specification.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS 8000 PERFUSION SYSTEM
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE - SARNS ROLLER PUMP
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5812013
MDR Text Key50771042
Report Number1828100-2016-00510
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K915183
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number16402
Device Catalogue Number16402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received09/26/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-