TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS 8000 PERFUSION SYSTEM; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE - SARNS ROLLER PUMP
|
Back to Search Results |
|
Model Number 16402 |
Device Problem
Naturally Worn (2988)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 06/28/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).This complaint is related to emdr #1828100-2015-00253.Per the srt, the indicator dial will be replaced with new part and returned for evaluation.
|
|
Event Description
|
The service repair technician (srt) reported that during routine testing of the device at the service center, the roller pump's indicator dial silks-creen increment markings are worn off.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.The service repair technician (srt) replaced the indicator dial with new part.The srt performed preventive maintenance (pm) and verification/release test.The roller pump operated to the manufacturer's specification.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|