Catalog Number 7205000000
Device Problem Unintended Power Up (1162)
Patient Problem No Patient Involvement (2645)
Event Date 06/30/2016
Event Type  malfunction  
Manufacturer Narrative
Device not yet received.
 
Event Description
It was reported by that the trigger was sticking, which has the potential to cause run-on.No patient involvement and no surgical delay were reported with this event.Additional information has been requested from the user facility.
 
Manufacturer Narrative
The reported event was confirmed through the evaluation service process but no run-on was found.
 
Event Description
It was reported by that the trigger was sticking, which has the potential to cause run-on.No patient involvement and no surgical delay were reported with this event.Additional information has been requested from the user facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYSTEM 7 DUAL TRIGGER ROTARY
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key5830049
MDR Text Key50659599
Report Number0001811755-2016-01327
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number7205000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received09/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/18/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-