MAUDE Adverse Event Report: BAXTER HEALTHCARE CLEARLINK DUO-VENT CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CLEARLINK DUO-VENT CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8541
Device Problems Kinked (1339); No Flow (2991)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/01/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a clearlink system continu-flo solution set experienced no flow.This occurred during patient use/setup.When the tubing is inserted inside of the sigma pump through the flow line, the tubing set becomes crimped, not allowing fluid to flow.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).Device manufacture date: 04/06/2016-04/07/2016.The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Clear passage and pressure testing were performed and passed successfully with no issues relating to the reported condition.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK DUO-VENT CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE
600 mts oeste de entrada
cartago
CS 
Manufacturer (Section G)
BAXTER HEALTHCARE
600 mts oeste de entrada
cartago
CS  
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5830678
MDR Text Key50686203
Report Number1416980-2016-12325
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K961225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 08/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/05/2021
Device Catalogue Number2C8541
Device Lot NumberR16D05102
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/08/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-