COOK ENDOSCOPY HERCULES 3 STAGE WIREGUIDED BALLOON ESOPHAGEAL-PYLORIC-COLONIC; KNQ, DILATOR, ESOPHAGEAL Back to Search Results
Catalog Number HBD-W-15-16.5-18
Device Problems Hole In Material (1293); Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/30/2016
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: cook qid-1 inflation device.Investigation evaluation: our evaluation of the returned device confirmed that there was a leak in the balloon.The balloon was inflated with a 60cc syringe and an inflation handle.There were three streams of water observed exiting the balloon from pinholes at varying locations near the proximal end of the balloon.The catheter was observed and there were no bends, kinks, or breaks.No part of the device was missing.A product discrepancy or anomaly that could have contributed to this reported occurrence was not observed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusions: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.The additional information provided indicated the balloon dilator did not receive negative pressure prior to advancement through the endoscope.This is the most likely cause for the reported observation.The instructions for use direct the user: "to facilitate passage through the endoscope, apply negative pressure to the device." the application of negative pressure will aspirate all residual air from the balloon and ease endoscopic advancement.Negative pressure will also aid in balloon preservation and optimize balloon performance.A possible contributing factor to balloon damage is inadequate lubrication of the balloon with a lubricating agent.The instructions for use direct the user to "apply a lubricating agent to the balloon to facilitate passage through the endoscope accessory channel." this activity will aid in endoscopic advancement and balloon preservation.Damage to the balloon material can occur if the balloon is inflated prior to advancement through the endoscope or if the balloon is inflated while partially or fully inside the accessory channel of the endoscope.The instructions for use contain the following precaution: ¿do not pre-inflate the balloon.¿ the instructions for use direct the user to ensure the balloon is completely visualized and positioned before inflation.The instructions for use contain the following warning: ¿during dilation, do not inflate balloon beyond the maximum indicated inflation pressure.¿ over inflation can cause damage to the balloon dilator.Another possible contributing factor is using a compromised inflation device to inflate the balloon.If the pressure reading of the inflation device is inaccurate, this could contribute to over inflation, possibly resulting in damage to the balloon material.Prior to distribution, all hercules 3 stage wireguided balloon esophageal-pyloric-colonic are subjected to a visual examination to ensure device integrity.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.Additional comments regarding this report: based on the information provided, a cook representative has been directed to contact the medical facility involved in an effort to promote further education and understanding related to appropriate usage of this product.
 
Event Description
During a pneumatic dilation, the physician used a cook hercules 3 stage wireguided balloon esophageal-pyloric-colonic.The dilation balloon was placed into the desired postion through [by way of the] wire guide.[the physician went to] inflate the balloon and the tail end [proximal end] of the balloon was found to be leaking.[the physician] replaced [the faulty device] with a new device to finish the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HERCULES 3 STAGE WIREGUIDED BALLOON ESOPHAGEAL-PYLORIC-COLONIC
Type of Device
KNQ, DILATOR, ESOPHAGEAL
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key5832807
MDR Text Key50721542
Report Number1037905-2016-00272
Device Sequence Number1
Product Code KNQ
UDI-Device Identifier00827002517974
UDI-Public(01)00827002517974(17)190104(10)W3667657
Combination Product (y/n)N
Reporter Country CodeCN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 07/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/29/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHBD-W-15-16.5-18
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Device Age6 MO
Event Location Hospital
Date Manufacturer Received07/06/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/04/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OLYMPUS GIF XQ260 ENDOSCOPE
Patient Age40 YR
Patient Weight80
-
-