Two service representatives observed the "blood detected" alarm in the fault logs and also observed excess of condensation.
No blood was present in any tubing.
They verified calibration of the condensate removal module, the fan operation of crm the validated calibration of blood detector sensor.
The service representatives performed both, condensate removal procedures.
Significant amount of condensation noted in purge line during removal process.
The iabp was tested to factory specification.
It functioned normally and was returned to the customer.
|