(b)(4).Upon completion of the investigation it was noted that the images were taken of the ¿as received¿ valve.The position of the cam when valve was received was 70mmh2o.The valve was hydrated for 24 hours.The valve was visually inspected: no defects were noted.The valve was tested for programming.With programmer 82-3126 with serial number (b)(4), the valve passed the test.The valve was flushed, the valve passed the test no occlusion was noted.The valve was leak tested, no leaks were noted.The valve was reflux tested.The valve passed the test.The catheter was irrigated, no occlusion was noted.The valve was dried.The valve was then pressure tested, the valve passed the test.Review of the history device records confirmed the valve product code 82-3112 with lot crkcwb, conformed to the specifications when released to stock in 29th september 2014.No root cause could be determined, as the problem reported by the customer could not be duplicated.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
|