It was reported that a patient had high impedance on a diagnostic test.The patient also reported an increase in seizures, even though it was noted that the patient's seizure rate was never completely controlled.A chest x-ray was performed, but due to the patient's body habitus, the view was limited.No further relevant information has been received to date.No known surgical intervention has occurred to date.
|
The explanted lead was returned in two portions and an analysis was completed on the two lead portions.A lead coil break was identified approximately 2mm from the electrode bifurcation.Due to extensive pitting, the coil fracture type could not be identified.Low magnification scanning electron microscopy found characteristics typical of lead discontinuity.Two set screw marks were found on the connector pin, indicating that at one point in time, a good mechanical and electrical connection was present.Other than the observed lead discontinuity, the conditions of the returned lead portions were consistent with conditions typical of an explant surgery.The explanted generator was returned and product analysis was performed.The generator was found to pass all electrical and functional specifications.The analysis of the generator from the pa lab concluded that no abnormal performance or any other type of adverse condition was found.
|