Device Problem
Material Discolored (1170)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
|
Event Description
|
The reporter contacted animas on (b)(6) 2016 alleging a display (dim/fading/color spectrum) issue.There was no alleged adverse event associated with this complaint.This complaint is being reported because the issue may impact the user's ability to read some or all of the information on the screen which may result in over or under delivery.
|
|
Manufacturer Narrative
|
Follow-up #1: date of submission 09/01/2016: device evaluation: the device has been returned and evaluated by product analysis on 08/09/2016 with the following findings: on investigation, the display was confirmed to be dim/fading and discolored.Unrelated to the original complaint, the battery compartment was cracked at case seal to the left of the bumper.
|
|