Device Problem Material Discolored (1170)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event Description
The reporter contacted animas on (b)(6) 2016 alleging a display (dim/fading/color spectrum) issue.There was no alleged adverse event associated with this complaint.This complaint is being reported because the issue may impact the user's ability to read some or all of the information on the screen which may result in over or under delivery.
 
Manufacturer Narrative
Follow-up #1: date of submission 09/01/2016: device evaluation: the device has been returned and evaluated by product analysis on 08/09/2016 with the following findings: on investigation, the display was confirmed to be dim/fading and discolored.Unrelated to the original complaint, the battery compartment was cracked at case seal to the left of the bumper.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key5845682
MDR Text Key52426793
Report Number2531779-2016-18790
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Type of Report Initial,Followup
Report Date 07/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2016
Was the Report Sent to FDA? Yes
Device Age41 MO
Date Manufacturer Received07/12/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-