Device Problems
Premature Discharge of Battery (1057); Crack (1135); Battery Problem (2885); Moisture or Humidity Problem (2986)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging that battery life of the pump was less than expected.The reporter claimed that the battery compartment was cracked and there was moisture in the battery compartment.This complaint is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.There was no indication that the product caused or contributed to an adverse event.
|
|
Manufacturer Narrative
|
Follow-up #1: date of submission 10/04/2016.Device evaluation: the device has been returned and evaluated by product analysis on 09/09/2016 with the following findings: a review of the pump¿s black box revealed evidence of a short battery life in a form of a sudden voltage drop leading to a cs 128 alarm.The battery compartment was found to be cracked at the threads on the side.The returned battery cap was able to fit securely.There was moisture corrosion observed in the battery canister.A leak test failed due to battery compartment leak.The pump powered up correctly and the ez-prime steps were performed correctly.All currents draw were within specification.The ¿short battery life¿ complaint was unable to be duplicated.The pump¿s cover was removed and there was no moisture ingress or any intermittent condition found to the printed circuit board.
|
|
Search Alerts/Recalls
|