On (b)(6) 2016 patient had a 50cc sensation plus iab catheter placed in the cath lab.
On (b)(6) 2016 at approx.
0553 the balloon pump began to alarm "gas loss in circuit," at this time the respiratory therapist checked all the connections.
No blood was found at this time.
The balloon was auto filled, restarted and quickly alarmed again.
Upon inspection the catheter and helium line were found to have blood in them.
The pump was stopped and the catheter was clamped off to stop the backwards flow of blood.
The cardiologist came in to replace the damaged balloon.
When the second replacement balloon was inserted it was found to be damaged (torn).
The third balloon was placed and therapy restarted.
The patient was transferred back to cicu.
On (b)(6) 2016 at about 1235 the pump once again began to alarm "gas loss in circuit.
" it was quickly discovered that once again there was blood in the catheter and helium line.
The therapist clamped the catheter and helium line.
This catheter was removed by the physician.
It was decided that it would not be replaced, as the plan was to remove it that day anyway.
About an hour later, the patient coded and expired.
|