The explanted devices were received for product analysis on 09/28/2016.Analysis was completed on the returned lead portions 10/11/2016.During the visual analysis a suture was observed sewn through a tie down and outer silicone tubing approximately 266mm from the end of the silicone tubes.The suture did not appear to have penetrated the inner silicone tube.The incision mark and the suture sewn through the outer silicone tubing and the cut ends made during the explanted process, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.For the observed inner tubing fluid remnants, there was no obvious path for fluid ingress other than the cut ends that were made during the explanted process.The condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No other obvious anomalies were noted except for the suture sewn through the outer silicone tubing.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, and no discontinuities were identified.Analysis was completed for the returned generator 10/18/2016.The septum was not cored.The device output signal was monitored for more than 24-hrs, while the pulse generator was placed in a simulated body temperature environment.Results showed no signs of variation in the pulse generatoroutput signal and demonstrated that the device provided the expected level of output current for the entire monitoring period.The pulse generator diagnostics were as expected for the programmed parameters.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.The battery measured 2.983 volts and shows the intensified follow-up indicator was not set.The downloaded data revealed that 42.695% of the battery had been consumed.Measured battery voltage and consumed capacity parameters were as expected.There were no performance or any other type of adverse conditions found with the pulse generator.
|