• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SND. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SND. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problems Corroded (1131); Low Battery (2584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/15/2016
Event Type  malfunction  
Event Description
It was reported that the external pulse generator indicated battery depletion despite having new batteries and that the batteries appeared to have been corroded by the device.The generator has not been returned for service.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: analysis confirmed the customer complaint of "battery depletion despite new batteries and batteries appear to have been corroded by the device", and noted that the battery contacts were corroded.At incoming testing the device passed all tests.The device was cleaned, checked and calibrated and then passed all tests on the automated test system.This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that the product was returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SND. BHD
bayan lepas free industr zone
bayan lepas
MY 
Manufacturer (Section G)
PLEXUS MANUFACTURING SND. BHD
bayan lepas free industr zone
bayan lepas
MY  
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5859679
MDR Text Key51773850
Report Number3004593495-2016-00325
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeEI
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 06/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received10/04/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/27/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1356-2015
Patient Sequence Number1
-
-