Model Number 2090 |
Device Problems
Failure to Power Up (1476); Sparking (2595); Noise, Audible (3273)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 06/22/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product event summary: it was confirmed that the programmer would not turn on, and there were popping and clicking sounds when attempting to power on.The programmer was found to have a faulty power supply and a loose power supply switch.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the programmer would not turn on.When an attempt was made to power on, there were sparks in the area of where the power cord plugs in, and an electrical popping and clicking sound could be heard.The programmer has been returned to service.There was no patient involvement.
|
|