• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION MAXPLUS® EXTENSION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION MAXPLUS® EXTENSION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number MP9209
Device Problem Fluid/Blood Leak (1250)
Patient Problem Hypoglycemia (1912)
Event Date 07/20/2016
Event Type  Injury  
Manufacturer Narrative
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
 
Event Description
The customer reported that an extension set that was used for an infusion of lipids and d12 at 5.3 ml/hr leaked at the filter.The patient's blood glucose was less than 30 for 4 hours despite d12 boluses, however one hour after changing the filter his blood glucose increased to 127.
 
Manufacturer Narrative
The customer¿s report of a filter leak was confirmed.Functional testing found that the filter was slightly misaligned within the body and there was a leak between the seams.The root cause of the leak was determined to be light seals on the filter which occurred due to build up on the heat stake during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXPLUS® EXTENSION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
stephen bilello
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key5870102
MDR Text Key51921647
Report Number9616066-2016-01057
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051499
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 07/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/01/2020
Device Model NumberMP9209
Device Catalogue NumberMP9209
Device Lot Number15086404
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/09/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight2
-
-