• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS PLATINUM 1 SERIES; SURGICAL ADJUNCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS PLATINUM 1 SERIES; SURGICAL ADJUNCTS Back to Search Results
Model Number 1MTEC30
Device Problems Bent (1059); Difficult or Delayed Positioning (1157)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/14/2016
Event Type  malfunction  
Manufacturer Narrative
The cartridge was returned to the manufacturer for evaluation.Visual inspection at 10x microscope magnification showed evidence of viscoelastic ovd (ophthalmic viscosurgical device) residues inside the cartridge.A deformed tip and a crack were observed at the cartridge tip.There was a shape hole at the tip section.The customer's reported complaint was confirmed.The manufacturing records could not be reviewed as the lot number was not provided.The directions for use (dfu) were reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue was verified.All pertinent information available to abbott medical optics has been submitted.
 
Event Description
It was reported that during implantation, the surgeon had difficulty injecting the intraocular lens (iol) into the patient's eye.When the cartridge was removed (from the eye) it was noticed that it was bent at the tip and a haptic had been sheared off the lens.The lens was then removed from the eye and a piece of the cartridge and the haptic were retrieved.A new lens was then implanted without difficulty.No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINUM 1 SERIES
Type of Device
SURGICAL ADJUNCTS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
Manufacturer (Section G)
ABBOTT MEDICAL OPTICS INC.
road 402 north, km 4.2
anasco industrial park, pob 14
anasco PR 00610
Manufacturer Contact
pam mcclain
1700 east st. andrew place
santa ana, CA 92705
7142478243
MDR Report Key5877805
MDR Text Key52609263
Report Number2648035-2016-01257
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081545
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Physician
Type of Report Initial
Report Date 08/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model Number1MTEC30
Device Catalogue Number1MTEC30
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/19/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-