MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON MB PATELLA PA A38; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON MB PATELLA PA A38; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5554L381
Device Problems Break (1069); Fracture (1260); Material Integrity Problem (2978); Insufficient Information (3190)
Patient Problems Pain (1994); Injury (2348)
Event Date 07/18/2016
Event Type  Injury  
Manufacturer Narrative
An evaluation of the device cannot be performed as the device was not returned to the manufacturer due to hospital policy.Additional information has been requested.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.Not returned to the manufacturer.
 
Event Description
Patient complaint of knee pain.
 
Manufacturer Narrative
An event regarding fracture involving a triathlon mb patellar component was reported.The event was confirmed.Method & results: device evaluation and results: a photograph of the articulating surface of the poly section of the explanted mb patellar component was provided.The poly component is fractured with a segment of the device missing.Medical records received and evaluation: clinician review of the x-rays and medical records provided indicated that complex mal-alignment and malposition of devices (baseplate and femoral component certainly) has contributed to instability in the knee with pain while also contributing to overload in the patello-femoral joint with patellar device damage.Device history review: dhr review for the reported lot determined that the device was manufactured and packed to specification.Complaint history review: complaint history review confirmed that there have been no other similar events for the reported lot.Conclusion: clinician review of the x-rays and medical records provided indicated that complex mal-alignment and malposition of devices (baseplate and femoral component certainly) has contributed to instability in the knee with pain while also contributing to overload in the patella femoral joint with patellar device damage.Further information such as device evaluation is needed to complete the investigation for determining a root cause.No further investigation is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
Patient complaint of knee pain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON MB PATELLA PA A38
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5878293
MDR Text Key52174845
Report Number0002249697-2016-02602
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/30/2015
Device Catalogue Number5554L381
Device Lot NumberSNLLJ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/22/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/04/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age76 YR
Patient Weight95
-
-