Model Number 4FC12 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Death (1802); Pericardial Effusion (3271)
|
Event Date 06/23/2016 |
Event Type
Death
|
Event Description
|
It was reported that post cryo ablation procedure the patient developed a pericardial effusion.The patient had a pericardiocentesis and fluid continued to build and the patient was taken to the operating room for a repair for the effusion.It was additionally reported that the patient died eleven days post procedure.Follow up occurred, but no further information has been obtained at this time.
|
|
Manufacturer Narrative
|
Product event summary: the data files were returned and analyzed.The data files did not show system notices or issues for the date of the event.There was no indication of a product malfunction, and no product was returned for investigation.A known clinical issue was encountered during the procedure.
|
|
Manufacturer Narrative
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|