The investigation has determined that higher than expected vitros tsh results were obtained from two levels of non-vitros quality control fluid using vitros tsh reagent.A definitive assignable cause could not be determined.There was no evidence to suggest a vitros 5600 integrated system malfunction.Unexpected instrument performance is not a likely contributor to the events, however it cannot be entirely ruled out as a contributing factor as precision testing to assess instrument performance was not undertaken.Based on historical vitros tsh reagent lot 5198 quality control results, there is no indication a reagent related issue contributed to the event.
|