Model Number 302-20 |
Device Problem
High impedance (1291)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/26/2016 |
Event Type
malfunction
|
Event Description
|
It was reported that a patient had high impedance identified by his physician.The physician ran multiple system diagnostic tests that gave high impedance results.The patient had a full replacement on (b)(6) 2016.The lead and generator were both returned to the manufacturer and are both currently undergoing product analysis.After replacement, the impedance values were found to be within normal limits.
|
|
Manufacturer Narrative
|
Supplemental mdr #1 inadvertently did not include information regarding pitting that was identified in the lead analysis.
|
|
Event Description
|
The abraded openings identified in analysis of the lead most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.Pitting and corrosion were identified on the coil surface of the lead.
|
|
Event Description
|
Product analysis on the returned lead portions was completed on 08/29/2016.A portion of one of the inner silicone tubes and quadfilar coils past the electrode bifurcation was not returned for analysis and therefore an evaluation could not be performed on those sections of the lead.The lead was returned in 6 portions.Visual analysis of the lead identified two coil breaks on one portion of the returned quadfilar coil.Scanning electron microscopy was performed on the two coil breaks, with one having extensive pitting which prevented identification of the coil fracture type, and the other broken coil end having evidence of a stress induced fracture with mechanical damage.Visual analysis identified another coil break on another portion of the lead, proximal to the anchor tether.Scanning electron microscopy identified the area as having evidence of a stress induced fracture with mechanical damage and no pitting.The presence of metal pitting suggested that stimulation was present for a certain period of time.Abraded openings were identified on the outer and silicone tubing in multiple locations.A half set screw mark was identified on the connector pin, indicating the lead had not been fully inserted into the generator cavity at one time.An additional set of screw marks were found on the lead connector pin indicating that at one point in time a good mechanical and electrical connection was present.No other anomalies were identified in review of the returned lead.The explanted generator was also returned and had product analysis completed on 09/12/2016.The generator confirmed proper functionality in its ability to measure impedance values as well as in its ability to provide appropriate output current.No abnormalities were identified for the generator.
|
|
Search Alerts/Recalls
|