It was reported by the company representative that the patient had come in for a vns replacement surgery.Upon interrogation of the device, the settings were found to be indicative of a faulted diagnostic test.It was explained to the surgeon that the settings found during interrogation were indicative of a faulted diagnostic test.The neurologist's physician's assistant was notified and stated the last time the checked the device was (b)(6) 2016 and he was able to provide the intended settings.The physician's assistant noted that when performing diagnostics on that date, the patient had a weird tightening sensation in her chest, so the diagnostics were interrupted.Due to the patient's tightness, they were unable to perform a final interrogation to verify the settings.It was explained to the physicians at that time that the patient would have been programmed at the unintended settings since (b)(6) 2016 until the day of surgery (2 weeks).After the implant was complete, the surgeon decided to program the patient back to the intended settings.
|