MAUDE Adverse Event Report: AMEDA, INC. PURELY YOURS; ELECTRIC BREAST PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMEDA, INC. PURELY YOURS; ELECTRIC BREAST PUMP Back to Search Results
Model Number 24502082
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/03/2016
Event Type  malfunction  
Manufacturer Narrative
The returned product was tested per ameda engineering protocol to determine whether the allegation of leaking fluid could be confirmed.The returned product was assessed for indications of malfunction or thermal event.The returned product was assessed for functionality and met functional specifications.Dark grey substance, e.G battery acid, was observed inside the battery compartment.Internally, no signs of foreign substance, burning, melting or charring were observed.The returned pump contained damaged batteries.The middle battery on top of the row of batteries was installed backwards.The battery fluid leakage was caused from the misplacement of the middle battery.
 
Event Description
Customer contacted ameda, inc.On (b)(6) 2016 to report an incident while pumping this morning using batteries to power on her purely yours breast pump.She states hearing a loud pop from inside the pump base during pumping and the pump stopped functioning.Customer looked inside the battery compartment and found several batteries exploded, having emptied black fluid inside the compartment.She reports no contact with this fluid and it stayed confined to the battery compartment.She denies injury, burn or property damage.A replacement purely yours pump base has been sent to customer to resolve this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PURELY YOURS
Type of Device
ELECTRIC BREAST PUMP
Manufacturer (Section D)
AMEDA, INC.
485 half day rd.
suite 320
buffalo grove IL 60089
Manufacturer Contact
linda zager
485 half day rd.
suite 320
buffalo grove, IL 60089
8479642620
MDR Report Key5919405
MDR Text Key53637467
Report Number3009974348-2015-00215
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973501
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Remedial Action Replace
Type of Report Initial
Report Date 08/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number24502082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received08/03/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age24 YR
-
-