• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER Back to Search Results
Model Number MERGE HEMODYNAMICS 10.0.1
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Failure to Transmit Record (1521); Device Issue (2379)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/08/2016
Event Type  malfunction  
Manufacturer Narrative
The customer's reported problem is currently under investigation by merge healthcare.When more information becomes available, a supplemental report will be submitted.
 
Event Description
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2016, a customer reported to merge healthcare that the hemo monitor shutdown during a procedure that resulted in no vitals, the pdm (patient data module) was buzzing, and the hemo monitor was black.With merge hemo not presenting physiological data during treatment, there is a potential for patient harm due to delay in treatment.However, information reported by the customer revealed that the procedure was completed successfully using external monitoring.Reference complaint number (b)(4).
 
Manufacturer Narrative
Merge healthcare conducted an internal quality investigation to address the issue reported in recall 2183926-02/15/2016-031-c, fda recall number z-0665-2017.Merge healthcare received reports of the hemo monitor application unexpectedly stopping to display and update patient data.When this issue occurs, the hemodynamics system is no longer capturing patient data.For customers who experience this issue, it is recommended that the hemo monitor pc is power cycled.Once the system is powered up, the hemo application should restart with normal functionality and will once again display, update and record patient data.The restarting of the hemo monitor pc may result 0630in a delay of up to two minutes while the system reboots.The investigation and troubleshooting activities conducted by merge healthcare found that the issue occurred due to an error when interfacing with a specific location within a schiller pdm (patient data module) file.Merge has validated and released a firmware fix for this issue.This fix is incorporated into the software upgrade of merge hemo 9.40.3 patch 1 (or later), or merge hemo 10.0.3 patch 1 (or later).The correction has been verified to be effective as is evidenced through the large reduction of customer complaints concerning this issue.Revised information contained in this supplemental report includes the following: h6 - evaluation codes: methods: 22 - software evaluation.Results: 110 - design error [the device or component had faulty (incomplete or incorrect) software design].Conclusions code: 12 - design deficiency [the device problem was traced back to the design specifications (e.G.In the requirements, testing processes, hazard analysis, implementation strategy].H10 - indication of additional manufacturer information is contained in this follow-up report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
MDR Report Key5938385
MDR Text Key54282957
Report Number2183926-2016-00717
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Remedial Action Recall
Type of Report Initial,Followup
Report Date 08/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMODYNAMICS 10.0.1
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/30/2017
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-0665-2017
Patient Sequence Number1
-
-