• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: 3M BAIR HUGGER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

3M BAIR HUGGER Back to Search Results
Device Problem Contamination (1120)
Patient Problem Unspecified Infection (1930)
Event Date 02/23/2010
Event Type  Injury  
Event Description
This report concerns injury to (b)(6) of (b)(6).On or about (b)(6), 2010, mr.(b)(6) underwent surgery during which bair hugger was used at (b)(6) by dr.(b)(6).It is believed that contaminants introduced into his open surgical wound by bair hugger resulted in a periprosthetic joint infection.He has undergone treatment(s) including, incision and drainage and multiple staged procedures to treat an infection beginning in (b)(6) 2012, at the (b)(6) by dr.(b)(6).Mr.(b)(6), his experts, and lawyers believe that the infection was caused by contamination of the sterile surgical field by the waste heat of the bair hugger forced-air warming system ("faw") used during his surgery.3m corporation, the mfr of bair hugger faw was notified in writing of the injury by litigation (case (b)(6) filed in the district of (b)(6) on (b)(6) 2016).Fda guidance dated july 9, 2013 provides that such legal documents constitute a "complaint" requiring the mfr to report to the fda.3m has failed to report this injury within the mandatory 30 days.Mr.(b)(6) and his lawyers assert that the approx 1000 watts of waste heat from bair hugger faw escapes from below the surgical drapes near the floor.It warms the contaminated air resident near the floor and forms into convection currents of rising contaminated warm air.The warm air rises alongside the surgical table, penetrating operating room ventilation airflow and ending up in the sterile surgical field.This phenomenon has been proved in at least six (6) studies published in peer-reviewed medical journals.Other studies have shown that the concentration of airborne contaminants above the wound positively correlates with the concentration of contaminants in the wound, and also positively correlates with the periprosthetic joint infection ("pji") rate.This leads to the inevitable conclusion that the waste heat from bair hugger faw must be increasing the risk of pji.At least one large outcome study has positively linked this rising waste faw heat to the majority of pji's after total joint replacement surgery.In contrast, there are no outcome studies showing that bair hugger faw is safe in implant surgery.In a separate issue, mr.(b)(6) asserts that 3m has willfully violated the terms of the bair hugger 510 (k).When 510 (k) #12345676 was filed in 2004, the mfr committed to a "hepa" quality inlet filter (99.97%) or greater filtration efficiency).In fact, the inlet filter of the bair hugger model 505 blower was never hepa and has now been reduced to 62% efficiency without notifying the fda.The result of the poor inlet filtration is that the internal airflow pathways of nearly all bair hugger blowers and hoses are contaminated with growing bacterial colonies.This has been shown in three studies published in peer-reviewed medical journals.The high-velocity airflow over the bacterial colonies has been shown to aerosolize bacteria out of the hose and into the operating room air.Since it is impossible to disinfect the internal airflow pathway of the bair hugger 505 blower, a hepa filter should be added to the outlet hose of these blowers to prevent the contamination of the operating room with the bacteria growing inside the blower and hose.Given the catastrophic nature of pji's, it is only prudent for the fda to err on the side of caution.Considering that there are several air-free and waste heat-free pt warming devices on the market today, any product that increases the risk of a pji should not be allowed in orthopedic surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BAIR HUGGER
Type of Device
BAIR HUGGER
Manufacturer (Section D)
3M
MDR Report Key5942613
MDR Text Key54463136
Report NumberMW5064661
Device Sequence Number1
Product Code DWJ
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation No Information
Type of Report Initial
Report Date 09/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention; Disability;
-
-